SA Configuration: portable with handle
AP Configuration: panel mount with flange
Ranges:
0-1000 ppmv
0-100 ppmv
0-10 ppmv
Atmosphere: Nitrogen
Input pressure range: 0.3 to 1.0 psig
Gas Sample Flow: 0.5 to 3.0 SCFH (limits)
Standard (primary) Output: Analog, 0-10Vdc. The 0-10Vdc analog output is directly related to the range
selected for display on the front panel. For example, if 0-100 ppm range is selected for display, the 0-10 Vdc out put
corresponds directly to 0-100 ppmv Maximum load resistance for this output is 750 ohm.
Optional Output: Analog, 4-20 mA. Completely isolated from any other output of the instrument. The 4-20
mA signal corresponds directly to the range selected for display on the front panel. Maximum load resistance for this output
is 750 ohm.
Alarms, Lo: Normally Closed (NC) and Normally Open (NO).
Form C contacts are provided when the instrument is equipped with a single point or dual point alarm relay. Contact ratings
are 3A, 30 Vdc and 5A, 115 or 230 Vac.
Alarms, High: Normally Closed (NC) and Normally Open (NO)
Form C contacts are provided when the instrument is equipped with a single point or dual point alarm relay. Contact ratings
are 3A, 30 Vdc and 5A, 115 or 230 Vac.
Switched AC: This signal is the input AC that is switched to the output connector whenever power to the
instrument is turned on. Its purpose is to provide power to an auxiliary data recorder. It is fused by the main instrument
fuse and should only be used for energizing a low power data recorder (<10 watts).
Internal Pump: verify
Gas Inlet: 1/8 inch compression (accepts 1/8 OD tube).
Gas Outlet: 1/8 inch compression (Connection to vent sample gas, not required, accepts 1/8 OD tube).
Voltage: 115 Vac (1 A max)
Case Size: 10.02" W x 11.02 H" x 6.06" D
Panel cutout: 8.00" W x 10.54 H" (SA-portable)
Depth in front of panel flange: 0.88"
Dept behind panel flange: 5.18"
Weight: verify